Posted Thu, 06 Oct 2022 16:47:43 GMT by Grudovik, Ilya
Trying to write a large contiguous buffer (8 000 000 samples) at a time:
Settings:
Digi V + [Freq=1MHz] + [Aperture=1mks] + [count = 8,000,000]

I see a pause in the data entry :(

In earlier versions of FW, there was no such error.
Posted Thu, 06 Oct 2022 16:48:31 GMT by Grudovik, Ilya
DMM7510.
Posted Wed, 23 Nov 2022 14:32:09 GMT by McKinney, Ty
Hello,
Apologies for the late response, I noticed this had been sitting here unanswered for a few weeks.
This is a bug that is known and we are working on a solution.
Can you tell me what firmware version you used previously where there was no issue? I would recommend downgrading to that version to fit your applications until a fix is presented.
If you can not find the firmware version that you need, where there was no pause in data entry, please open a case with our technical support center and we can work with you individually. If you make a case with us, please reference your TekTalk forum post so we can sort it out to the correct team.
Posted Wed, 23 Nov 2022 15:23:05 GMT by Grudovik, Ilya
Hi!
I'm very happy with the live answer :)

If the problem is known and it will be solved, then I will wait.

While I'm working with a smaller buffer. A larger buffer would be more convenient. But this is not a reason to roll back the firmware. This is an old firmware where there was no general trigger.BLOCK_MEASURE_DIGITIZE command.