Posted Fri, 25 Nov 2022 04:22:23 GMT by 王, 小奇
Hi  
I use c # to write a program to control DAQ6510 to collect data. Is there a sample program to switch the instrument to continuous trigger mode

thanks
Nickey
Posted Mon, 28 Nov 2022 15:25:22 GMT by A, Jake
Hello,

Continuous Trigger mode is only available via the front panel, not remote operation.
When remote programming, we would recommend developing a trigger model that will provide more control over the buffer and operational details of your test. You can use an infinite trigger, if you prefer, to continually record data in a similar fashion.
Continuous Trigger mode is designed for data reference as a bench instrument and is not compatible with programming or remote operation.